Nuvo Q

Specifications

Specification Model 500-003Q Model 900-003Q
Electrical Requirement 115 Volt – 60 Hz 220 Volt – 50/60 Hz
Flow Delivery Rate 0 to 5 LPM (1/2 Liter Increments)
Oxygen Concentration 0.5 to 4 liters per minute – 93% (+6.5% / -3%)
At 5 liters per minute – 90% (+6.5% / -3%)
Sound Level < 40 dBA typical
Power Consumption < 400 Watts average
Operating System Rapid Pressure Swing
Oxygen Monitoring System Available – Model 505-003Q Available – Model 905-003Q
Safety Alarms & Features Power Failure
Process Failure
High Pressure / Low Pressure
Low Oxygen Concentration (in OMS Model)
Current overload or line surge shutdown
Thermal Switch
40 psig Pressure Relief Valve
Low Battery Test
Filters Cabinet, Compressor Intake & Bacteria
Battery 9 Volt
Weight < 50 lbs or < 23 kg
Dimensions 14”W x 9”D x 23”H
394mm L x 396mm W x 706mm H
Operating Environment Ambient Temperature: 50°F to 100°F (10°C to 40°C)
Humidity: 15% to 95%, non-condensing
Storage Range Temperature: 0°F to 140°F (-20°C to 60°C)
Humidity: 15% to 95%, non-condensing